Ra Readings for Product Contact Surfaces. Benway, Chair, Swagelok Co.

Nemenoff, Habonim Industrial Valves Ltd. Kettermann, Rath Gibson K. This is usually achieved at a velocity of 5 fps 1. Where necessary, increases in height should be achieved by vertical risers see Fig.

The bore of weld neck flanges shall be the same as the I. If it is not practical to conduct the test with all interior equipment in place, dummy shafts and dip tubes may be used to check shadowing. Basic Components of a Seal.

ASME BPE Bioprocessing Equipment_图文_百度文库

Weld Joint Design and Preparation. The Standards Committee that approved asmw code or standard was balanced to assure that individuals from competent asmee concerned interests have had an opportunity to participate. Zinkowski, Vice Chair P. Peterman, High Purity Connections, Inc. Gravity then provides for a continuous solution sheeting over the lower surfaces. SD Redesignated as Fig. This Part describes the fittings made for use with nominal outside diameter O.

Process lines should be cleaned at a velocity to ensure turbulent flow in pipelines, and maintain a fully flooded line. Tagging nozzles on the backside of panels will help reduce the number of incorrect piping connections during field installation.

  IN THE GARDEN OF TEMPTATION CYNTHIA WICKLUND PDF

Such usage may be by an association or a society, or by a holder of a Symbol Stamp who may also use the facsimile in advertising to show that clearly specified items will carry the symbol.

MacDonald, United Association D. However, all the previously mentioned construction codes shall be satisfied including those instances where these codes are not referred to in the current BPE Standard e.

If requested by the end-user, the manufacturer shall supply a sectioned sample of the bend area.

In all cases sufficient annular space to allow access for CIP coverage shall be provided. Shaft Coupling Seal Arrangement. Hobick, Holland Applied Technologies C. The manufacturers of agitators and mixers shall verify the cleanability of their equipment as specified and agreed to with the end-user.

Schnell, Dupont Performance Elastomers D. Hutton, Chair, Arkema, Inc. Sisto, Purity Systems, Inc.

Items constructed in accordance with all of the applicable rules of ASME are identified with the official Symbol Stamp described in the governing code or standard. It is the responsibility of the end-user to establish complaince with applicable standards e. All encompassing lines including long runs with the exception of subheaders, manifolds, and nozzles may be sloped as defined in SD The break shall be at least 1 in.

The product contact surfaces shall be free of crevices, pockets, and other surface irregularities. The line size reduction can be made after the branch to the trap leg see Fig. Pembleton, Dupont Fluoropolymer R.

SD-7, illustrations c and d. Water used for the following rinse shall meet the requirements of SD Concentric and Eccentric Reducer. An organization holding authorization to perform various activities in accordance with the requirements of the codes and standards may state this capability in its advertising literature.

  CALCITRAN D3 BULA PDF

Fittings shall be so designed that there will not be any crevices or hard-to-clean areas around the gasketed joint. Technical documentation to support the design of equipment and verify conformance with sterility and cleanability criteria may include, but not be limited to, the following: The light fixture shall be sealed. Physical Break in Point-of-Use Piping.

Rau, Dockwelier AG B. Pure Steam a The clean steam distribution system shall have adequate provision to remove air during start-up and normal operations. It is the aim of the Society to maintain the standing of the Symbol Stamps for the benefit of the users, the enforcement jurisdictions, and the holders of the Stamps who comply with all requirements.

ASME BPE Guidelines for Pharmaceutical Equipment

Acceptance Criteria for Welds on Tube. Harrison, Harrison Electropolishing L. Klemp, Advance Fittings Corp. If a smaller annular space is provided, other methods of cleaning this space should be provided, such as holes in the dip tube that CIP the nozzle spacing.

The conformance of material should be explicitly stated e.